Manufactured in the U.S.A by IDM Tech, LLC for CryoPen, Inc.
Distributed by CryoPen, Inc
CryoPen
®
is a Registered Trademark of CryoPen, Inc. All Contents Copyright © All Rights Reserved
- 11 -
Originated Date: 02.25.2010
Revision Date: 5.1.2012
OM-CT2-01-R6
CLEANING:
The CryoPen
®
Cooling System: The main unit, except for the CryoPen
®
chilling wells, may be cleaned and disinfected as
needed using isopropyl alcohol or Reservoir Solution. CryoPen
®
chilling wells may be cleaned periodically as dust and
debris may collect in them to the point of making poor contact with the Pen Cores. To clean these chilling wells, use the
provided swabs. Use a blotting and circular motion to clean out chilling wells. Never place your fingers into the chilling
wells, due to the extreme cold. Do not submerge the unit into water; doing so will present an electrical hazard and destroy
the unit. It is important not to leave chilling wells open for an extended period of time. Swabbing the wells can be
completed quickly when rotating pens.
The CryoPen
®
Pen Cores: may be wiped cleaned with soap and water when the Pen Cores are at room temperature (70-75°
degrees Fahrenheit) or cleaned (with such standard solutions as isopropyl alcohol or CryoPen’s Reservoir Solution) as
needed. Do not submerge pen cores into liquid of any kind. Doing so will render them inoperable. CryoPen
®
Pen Cores
should never be used for a procedure without the appropriate tip in place.
Reservoir Solution: Should be filled to the maximum line at all times and should be completely changed weekly. Swab and
dry out tube completely before refilling. Refill with reservoir solution from 500ml. bottle. Note: The use of any other
solution other than that obtained from CryoPen, Inc. will void warranty.
DAILY MAINTENANCE:
Daily maintenance can be done while the unit is on.
Rotate pen cores from the holding wells to the chilling wells at the end of each day. During weekends, it is not
necessary to rotate pen cores. (Make sure all pen cores from holding wells are wiped dry and dipped before placing
into chilling wells.)
Reservoir tube should be checked daily and filled as needed to the maximum line.
Reservoir solution level in tube should be kept at the maximum level at all times. Remove Reservoir Cap if used
and discard old solution at the end of each week. Dry tube thoroughly using the supplied swab, replace tube into
unit. Refill the tube with supplied Reservoir Solution stored in the 500ml bottle. Open cap and dispense into
reservoir tube.
Reorder 500ml Reservoir Solution as needed.
Clean and disinfect system as needed by wiping down with standard isopropyl alcohol or Reservoir Solution.
Make sure unit and surrounding area is kept free from dust.
Periodically remove holding wells from unit and clean with warm soapy water and air dry.